Contents
Nitrogen–phosphorus detector
The nitrogen–phosphorus detector (NPD) is also known as thermionic specific detector (TSD) is a detector commonly used with gas chromatography, in which thermal energy is used to ionize an analyte. It is a type of flame thermionic detector (FTD), the other being the alkali flame-ionization detector (AFID also known as AFD). With this method, nitrogen and phosphorus can be selectively detected with a sensitivity that is 104 times greater than that for carbon.
NP-Mode
A concentration of hydrogen gas is used such that it is just below the minimum required for ignition. A rubidium or cesium bead, which is mounted over the nozzle, ignites the hydrogen (by acting catalytically), and forms a cold plasma. Excitation of the alkali metal results in ejection of electrons, which in turn are detected as a current flow between an anode and cathode in the chamber. As nitrogen or phosphorus analytes exit the column, they cause a reduction in the work function of the metal bead, resulting in an increase in current. Since the alkali metal bead is consumed over time, it must be replaced regularly.
This article is derived from Wikipedia and licensed under CC BY-SA 4.0. View the original article.
Wikipedia® is a registered trademark of the
Wikimedia Foundation, Inc.
Bliptext is not
affiliated with or endorsed by Wikipedia or the
Wikimedia Foundation.